The Big Instrument is the main instrument used in the CC Jarrold group. Several different experiments can be performed on this instrument including photoelectron spectroscopy, reactivity of anions, and resonant two-photon detachment. All of these experiments begin by generating met al suboxide ions in the source region. The most common experiment performed on this instrument is photoelectron spectroscopy, so the apparatus is described with respect to this experiment.
Note that the schematic of the apparatus includes a VMI region that the group hopes to install when funding becomes available.
A distribution of pure or binary metal oxide anions are generated using a pulsed molecular beam valve/ablation laser ion source, a simplified diagram of this source is shown to the right. Helium carrier gas is introduced using a pulsed (30 Hz) valve (beam valve 1) and approximately 8 mJ/pulse of the second harmonic (532 nm) of a Nd:YAG, is focused onto a rotating rod (not shown) or pressed metal powder source. The pellet source is made by pressing the metal powder at high (2-3 tons) pressure into a solid pellet. By using this method, we are easily able to make samples with mixtures of more than one metal (i.e. Mo-V, Mo-Nb, Mo-Al).
The ions are then skimmed and the anions are accelerated to 1 keV (green in figure) into a time-of-flight spectrometer (blue in figure). The ions pass through a 3-mm mass-defining slit into the detector region (red in figure) and collide with a dual microchannel plate detector assembly. Anions are photodetached 15 cm upstream of the ion detector using the second (532 nm, 2.33 eV), third (355 nm, 3.49 eV), or fourth (266 nm, 4.66 eV) harmonic output of a second Nd:YAG laser, timed to intersect a particular ion packet. There are many ways that photodetached electrons can be extracted and detected and they all have different advantages and disadvantages. The method used on this instrument is called time-of-flight. In this method, a small fraction of the photodetached electrons travel the length of a 1 m field-free drift tube situated perpendicular to both the ion and laser beam axes. Data collection is slower for this method compared to others such as a magnetic bottle, since only a very small fraction of the electrons are detected (opposed to 100% for a magnetic bottle). However, time-of-flight methods produce higher resolution data and angular information is not lost, so anisotropy parameters can be measured.
The drift times of the photoelectrons are recorded relative to the detachment laser pulse using a digitizing oscilloscope. The drift times are averaged over 1,000,000 laser shots and converted to electron kinetic energy.
The energies are calibrated using well-known PE spectra, such as O−, OH− and WO2−. Besides photoelectron spectroscopy, we also perform ion beam hole-burning experiments, reactivity studies, and resonant two-photon detachment (R2PD). The instrument is operated slightly different for these experiments.